To eliminate measurement error, the CIH20DL must be moved close enough so the
motor is the only object in the target area of its IRT (see below figure). For a motor
with an area of 1 ft
2
and using an IRT with a D:S ratio of 8:1, the optimum
measurement distance would be 8 ft.
For best results, the targeted area (spot) should fall within the target’s boundaries
Measuring a motor’s temperature from the
wrong (top) and right (bottom) distance
25