Calibration Method:
Turn off the meter. Keep the MODE (CAL) key pressed
and press key to turn on the power. When display
indicate symbol “CAL” to enter calibration Mode. The
indication reading should be 18.4% (±0.5). If need to
calibrate, please use screwdriver to calibrate from right-side
of meter.
After calibration, please turn off the meter to save .
Notice
The calibration reading will have deviations in
the operation ambient.
Notice
When calibration, please make sure the probe
suspended in midair and can’t contact any
object.
- 6 -
D. The LCD Display
- 3 -
Manual Power Off
Calibration Mode
Recording Mode
Moisture Value
No. of Calibration Table
Low battery
MAX Value
Hold Value
MIN Value
Recording value warns